TIDA-01614 Reference Design

Texas Instruments

Multiparameter front end reference design for vital signs patient monitor

Description

This reference design is for a multiparameter front end of a patient monitor that measures vital sign parameters like ECG: heart rate: SpO2 and respiration.it uses biosensing front end integrated circuits like the AFE4403 and ADS1292R devices: to measure these parameters. It also uses three TMP117 sensors to accurately measure skin temperature. This design can interface with the pace detection module to detect the pace pulse. The design also uses an isolated UART connection to transfer data to a computer. The entire front end subsytem runs on a rechargeable 3.7V li ion battery.

Features
  • Monitors ECG: Heart rate: SpO2: respiration rate and skin temperatureUses biosensing front end AFE4403 for SpO2 and heart rate measurement: and ADS1292R for ECG and respiration measurement. Supports up to three LEDs and three photodiodes with ambient subtraction to improve signal to noise ratio for SpO2 and heart rate measurement. Single lead ECG measurement with RLDSupports three 0.1 Celsius accurate sensors to measure the skin temperatureInterfaces to pace detection module(software configurable  cardiac pacemaker detection module reference design ) to enable pacemaker detectionEnables data transfer over isolated UART interfaceRuns on a one cell Li Ion rechargeable battery
Applications
  • Pulse oximeter
  • Electrocardiogram (ECG)
  • Multiparameter patient monitor
  • Medical sensor patches
Product Categories
  • Data converters